Buy me a coffee
Home › FDA recalls › ICU Medical Plum Duo Infusion Pump
CLASS II · ModerateFDA

ICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump

📅 Dec 15, 2025 · 📍 Nationwide
Recalling company
ICU Medical, Inc.
Product
ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
Reason
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 15, 2025
Reported
Jan 28, 2026
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Status
Ongoing
Lot / codes
400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0
Company location
Lake Forest, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1118-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1118-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.