Buy me a coffee
Home › FDA recalls › Medline medical procedure convenience kits labeled as: 1)
CLASS II · ModerateFDA

Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE

📅 Dec 16, 2025 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.
Reason
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 16, 2025
Reported
Jan 21, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
1457 units
Lot / codes
REF CDS980754T: UDI/DI 10198459121814 (EA) 40198459121815 (CS), Lot Number 25EMD795; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBR237; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBP591; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25DBM353; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25CBL184; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25BBL633; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25FBK918; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25CBJ496; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25FBF371; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25EBK493; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25DBH895; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25DBC272; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBU642; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBJ298; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25BBU206; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25FBA261; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25CBN230; REF DYNJ901829T
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1089-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1089-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.