Home › FDA recalls › Medline medical procedure convenience kits labeled as: 1)
CLASS II · ModerateFDA

Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS

📅 Dec 16, 2025 · 📍 Nationwide
Recalling company
Medline Industries, LP
Product
Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.
Reason
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 16, 2025
Reported
Jan 21, 2026
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
4536 units
Lot / codes
REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25FBK870; REF CDS860237AF: UDI/DI 10198459232626 (EA) 40198459232627 (CS), Lot Number 25EBD110; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25EBC253; REF CDS982128Q: UDI/DI 10198459008245 (EA) 40198459008246 (CS), Lot Number 25DBG393; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25FBQ303; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25EBD631; REF CDS984338M: UDI/DI 10198459148101 (EA) 40198459148102 (CS), Lot Number 25DBM322; REF CDS984640I: UDI/DI 10198459318955 (EA) 40198459318956 (CS), Lot Number 25DBT807; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25DMG099; REF DYNJ0525321O: UDI/DI 10198459236907 (EA) 40198459236908 (CS), Lot Number 25CMJ006; REF DYNJ57640F: UDI/DI 10198459011115 (EA) 40198459011116 (CS), Lot Number 25FDB347; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25EBK539; REF DYNJ62594D: UDI/DI 10198459049767 (EA) 40198459049768 (CS), Lot Number 25DBD145; REF DYNJ63392D: UDI/DI 10198459164507 (EA) 40198459164508 (CS), Lot Number 25BLA967; REF DYNJ68187B: UDI/DI 10198459382086 (EA) 40198459382087 (CS), Lot Number 25FMG440; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25EMH020; REF DYNJ82368A: UDI/DI 10198459258077 (EA) 40198459258078 (CS), Lot Number 25CMF917; REF DYNJ85261:
Company location
Northfield, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-1087-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1087-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.