Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS
- Recalling company
- Medline Industries, LP
- Product
- Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359
- Reason
- Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Dec 16, 2025
- Reported
- Jan 21, 2026
- Distribution
- US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 96 units
- Lot / codes
- 1) UDI/DI 10198459269196 (EA) 40198459269197 (CS), Lot Number 25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS), Lot Number 25FBA103.
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1085-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1085-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.