Buy me a coffee
Home › FDA recalls › Philips Azurion Systems R3.0. Includes the following label d
CLASS II · ModerateFDA

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding

📅 Dec 8, 2025 · 📍 Nationwide
Recalling company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
Reason
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 8, 2025
Reported
Jan 21, 2026
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Status
Ongoing
Quantity
990 units
Lot / codes
Software Version: R3.0. 1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29. 2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22. 3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72, (01)00884838116740(21
Company location
Best
Recall type
Voluntary: Firm initiated
Recall number
Z-1079-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1079-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.