Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License
- Recalling company
- Elekta, Inc.
- Product
- MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
- Reason
- Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Dec 10, 2025
- Reported
- Jan 21, 2026
- Distribution
- US Nationwide distribution in the state of TX.
- Status
- Ongoing
- Quantity
- 1
- Lot / codes
- MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071
- Company location
- Atlanta, GA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-1078-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1078-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.