REPLIGEN CORPORATION recalls MAVERICK¿ Measurement Module Spectrometer PN 820-01367
- Recalling company
- REPLIGEN CORPORATION
- Product
- MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Reason
- Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Dec 10, 2025
- Reported
- Jan 21, 2026
- Distribution
- U.S.
- Status
- Ongoing
- Quantity
- 33
- Lot / codes
- PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Company location
- Marlborough, MA
- Recall type
- FDA Mandated
- Recall number
- Z-1024-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-1024-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.