Buy me a coffee
Home › FDA recalls › MAVERICK¿ Measurement Module Spectrometer PN 820-01367
CLASS II · ModerateFDA

REPLIGEN CORPORATION recalls MAVERICK¿ Measurement Module Spectrometer PN 820-01367

📅 Dec 10, 2025 · 📍 See notice
Recalling company
REPLIGEN CORPORATION
Product
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Reason
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 10, 2025
Reported
Jan 21, 2026
Distribution
U.S.
Status
Ongoing
Quantity
33
Lot / codes
PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Company location
Marlborough, MA
Recall type
FDA Mandated
Recall number
Z-1024-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1024-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.