Buy me a coffee
Home › FDA recalls › BVI Leos Laser and Endoscopy System.
CLASS II · ModerateFDA

Beaver-Visitec International, Inc. recalls BVI Leos Laser and Endoscopy System.

📅 Dec 5, 2025 · 📍 Nationwide
Recalling company
Beaver-Visitec International, Inc.
Product
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Reason
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 5, 2025
Reported
Jan 7, 2026
Distribution
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Status
Ongoing
Quantity
5 units
Lot / codes
Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
Company location
Waltham, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-1008-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-1008-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.