Buy me a coffee
Home › FDA recalls › Brand Name: Affixus Antegrade Femoral Nailing System Product
CLASS II · ModerateFDA

Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femor

📅 Dec 2, 2025 · 📍 Nationwide
Recalling company
Zimmer, Inc.
Product
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Right, Trochanteric Nail Component: N/A
Reason
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Dec 2, 2025
Reported
Dec 31, 2025
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
Status
Ongoing
Quantity
22
Lot / codes
Lot Code: Model No 815509420 ; UDI-DI (01)00887868143571(17)340920(10)66717556 ; Lot Number 66717556 Model No 815509420 ; UDI-DI (01)00887868143571(17)340920(10)66884566 ; Lot Number 66884566
Company location
Warsaw, IN
Recall type
Voluntary: Firm initiated
Recall number
Z-0972-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0972-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.