Buy me a coffee
Home › FDA recalls › LifeShield Drug Library Management
CLASS II · ModerateFDA

ICU Medical, Inc. recalls LifeShield Drug Library Management

📅 Nov 24, 2025 · 📍 Nationwide
Recalling company
ICU Medical, Inc.
Product
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
Reason
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Nov 24, 2025
Reported
Dec 31, 2025
Distribution
US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
Status
Ongoing
Quantity
12 customers
Lot / codes
LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227
Company location
Lake Forest, IL
Recall type
Voluntary: Firm initiated
Recall number
Z-0966-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0966-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.