Home › FDA recalls › The Stryker CranialMask Tracker is a single-use device consi
CLASS II · ModerateFDA

Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed ci

📅 Nov 12, 2025 · 📍 Nationwide
Recalling company
Howmedica Osteonics Corp.
Product
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Reason
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Nov 12, 2025
Reported
Dec 31, 2025
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Status
Ongoing
Quantity
320 units
Lot / codes
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Company location
Mahwah, NJ
Recall type
Voluntary: Firm initiated
Recall number
Z-0963-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0963-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.