Medtronic MiniMed, Inc. recalls InPen App
- Recalling company
- Medtronic MiniMed, Inc.
- Product
- InPen App, Model/CFN Number: MMT-8061 (Android Users)
- Reason
- When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Nov 13, 2025
- Reported
- Dec 24, 2025
- Distribution
- US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
- Status
- Ongoing
- Quantity
- 8251
- Lot / codes
- Software Versions: 7.5.0, 7.5.1, and 8.0.0. UDI-DI: 00763000974596
- Company location
- Northridge, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0958-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0958-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.