Buy me a coffee
Home › FDA recalls › Brand name: AGFA Digital Radiography X- Ray system DR 800 wi
CLASS II · ModerateFDA

AGFA Healthcare Corp. recalls Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name

📅 Nov 18, 2025 · 📍 Nationwide
Recalling company
AGFA Healthcare Corp.
Product
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Reason
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Nov 18, 2025
Reported
Dec 31, 2025
Distribution
U.S. Nationwide distribution.
Status
Ongoing
Quantity
35
Lot / codes
N/A
Company location
Greenville, SC
Recall type
FDA Mandated
Recall number
Z-0923-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0923-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.