Mindray DS USA, Inc. dba Mindray North America recalls Hardware configuration of the BeneVision Central Monitoring System
- Recalling company
- Mindray DS USA, Inc. dba Mindray North America
- Product
- Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1 HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)
- Reason
- When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Nov 18, 2025
- Reported
- Dec 24, 2025
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Status
- Ongoing
- Quantity
- 906 units
- Lot / codes
- Audio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP. 1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902 2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4
- Company location
- Mahwah, NJ
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0920-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0920-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.