Buy me a coffee
Home › FDA recalls › GE Healthcare Carestation anesthesia system
CLASS I · SeriousFDA

GE Medical Systems China Co., Ltd. recalls GE Healthcare Carestation anesthesia system

📅 Nov 14, 2025 · 📍 Nationwide
Recalling company
GE Medical Systems China Co., Ltd.
Product
GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 610 A1, REF 1012-9620-222; Carestation 620 A1 REF 1012-9620-000; Carestation 620 A1, REF 1012-9620-200; Carestation 620 A1, REF 1012-9620-202; Carestation 620 SE A1, REF 1012-9620-212; Carestation 630 A1, REF 1012-9650-222; Carestation 650 A1, REF 1012-9650-000; Carestation 650 A1, REF 1012-9650-200; Carestation 650 A1, REF 1012-9650-202; Carestation 650 SE A1, REF 1012-9650-212; Carestation 650c A1, REF 1012-9655-202; Anesthesia Gas Machine Products for which affected PMB can be used as a spare part: Carestation 620 A2, REF 1012-9620-002 00840682124546 Carestation 620 SE A2, REF 1012-9620-012 00195278569677 Carestation 650 A2, REF 1012-9650-002 00840682124560 Carestation 650 SE A2, REF 1012-9650-012 00195278569684 Carestation 650c A1, REF 1012-9655-000 00840682103954 Carestation 650c A2, REF 1012-9655-002 00840682124539 Carestation 650c A1, REF 1012-9655-200 00195278439543
Reason
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Nov 14, 2025
Reported
Dec 24, 2025
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.
Status
Ongoing
Quantity
1277 units
Lot / codes
REF 1012-9620-222: GTIN 00195278626301; REF 1012-9620-000: GTIN 00840682103985; REF 1012-9620-200: GTIN 00195278439536; REF 1012-9620-202: GTIN 00195278626158; REF 1012-9620-212: GTIN 00195278626561; REF 1012-9650-222: GTIN 00195278626592; REF 1012-9650-000: GTIN 00840682103947; REF 1012-9650-200: GTIN 00195278439529; REF 1012-9650-202: GTIN 00195278626585; REF 1012-9650-212: GTIN 00195278625687; REF 1012-9655-202: GTIN 00195278625953; REF 1012-9620-002: GTIN 00840682124546; REF 1012-9620-012: GTIN 00195278569677; REF 1012-9650-002: GTIN 00840682124560; REF 1012-9650-012: GTIN 00195278569684; REF 1012-9655-000: GTIN 00840682103954; REF 1012-9655-002: GTIN 00840682124539; REF 1012-9655-200: GTIN 00195278439543.
Company location
Wuxi
Recall type
Voluntary: Firm initiated
Recall number
Z-0913-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0913-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.