Home › FDA recalls › AMSCO 7053HP Washer/Disinfector
CLASS II · ModerateFDA

Steris Corporation recalls AMSCO 7053HP Washer/Disinfector

📅 Nov 19, 2025 · 📍 Nationwide
Recalling company
Steris Corporation
Product
AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
Reason
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Nov 19, 2025
Reported
Dec 17, 2025
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
Status
Ongoing
Quantity
22
Lot / codes
Lot Code: UDI: 00724995176662, 00724995176679, 00724995176655 Serial numbers of distributed units: 3619725008, 3619725011, 3619725013, 3621625010, 3621625011, 3621625013, 3621625014, 3621625015, 3621925001, 3621925002, 3621925005, 3621925010, 3622425005, 3622725005, 3622725011, 3622725012, 3623125006, 3623425016, 3623825008, 3624025002, 3624025008, 3624525007
Company location
Mentor, OH
Recall type
Voluntary: Firm initiated
Recall number
Z-0910-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0910-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.