Bard Peripheral Vascular Inc recalls Safe-T-Centesis 6 Fr Catheter Drainage Tray
- Recalling company
- Bard Peripheral Vascular Inc
- Product
- Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
- Reason
- Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Nov 6, 2025
- Reported
- Dec 17, 2025
- Distribution
- US Nationwide distribution in the states of CA, MN, IN, MI, LA.
- Status
- Ongoing
- Quantity
- 1240
- Lot / codes
- UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
- Company location
- Tempe, AZ
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0887-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0887-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.