Philips North America Llc recalls Incisive CT
- Recalling company
- Philips North America Llc
- Product
- Incisive CT; Software Version Number: 4.5, 5.0, 5.1;
- Reason
- If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Oct 29, 2025
- Reported
- Dec 10, 2025
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
- Status
- Ongoing
- Quantity
- 2341 units
- Lot / codes
- 1) Model Number: 728143; UDI-DI: 00884838085015; Serial Numbers: 500092, 500202, 500309, 500376, 500574, 500265, 500264, 510051, 500034, 500544, 500543, 500548, 500046, 500168, 500064, 500110, 500111, 500102, 500145, 500178, 500255, 500217, 500211, 500216, 500225, 33009, 500242, 500252, 500222, 500282, 33005, 500301, 33001, 33035, 33039, 33002, 33003, 33007, 33013, 33047, 33011, 33008, 33015, 33020, 33010, 33046, 33017, 33019, 33049, 33016, 33024, 33006, 33025, 33034, 33041, 33038, 33031, 33042, 33033, 33037, 33045, 33012, 500032, 500073, 500138, 500308, 33023, 500443, 500410, 500412, 500557, 500038, 500069, 500332, 500397, 510056, 500091, 500135, 510028, 510029, 500591, 500123, 500087, 500136, 500103, 500116, 500105, 500056, 500125, 500041, 500089, 500200, 500152, 500219, 500256, 500213, 500307, 510001, 500233, 500371, 500370, 500377, 510030, 510032, 510031, 500360, 510005, 510020, 500329, 500361, 500436, 500457, 500406, 500465, 500525, 500535, 510027, 500523, 33043, 500028, 500024, 500122, 500173, 510095, 500374, 500003, 500020, 500052, 500075, 500118, 500149, 500150, 500134, 500146, 500218, 500223, 500232, 500299, 500311, 500394, 500435, 500463, 33028, 500480, 500078, 500157, 500153, 500229, 500276, 500310, 500300, 500318, 500336, 500347, 500423, 500372, 500357, 500597, 500351, 500363, 500414, 500390, 500448, 500598, 500399, 500405, 500428, 500391, 500524, 500498, 500422, 500526, 500542, 500541, 500565, 500575, 500578, 500592, 500585, 500593, 500586, 500604, 500581, 500577
- Company location
- Cambridge, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0881-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0881-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.