Magellan Diagnostics, Inc. recalls The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to
- Recalling company
- Magellan Diagnostics, Inc.
- Product
- The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
- Reason
- Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Feb 23, 2024
- Reported
- Dec 3, 2025
- Distribution
- US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 1,774 units
- Lot / codes
- Product Name: LeadCare¿ II Blood Lead Test Kit; Catalog Number: 70-6762; Impacted Lot/Sublot Numbers: 2333M, 2333M-01, 2333M-02, 2333M-03, 2333M-04, 2333M-05, 2333M-06, 2333M-07, 2333M-08, 2333M-09, 2333M-10, 2333M-11; Device Identifier: (01)00850355006000;
- Company location
- North Billerica, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0584-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0584-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.