Buy me a coffee
Home › FDA recalls › REF: CC286297 or BS737162 Automated Digital Cell Morphology
CLASS II · ModerateFDA

Cellavision AB recalls REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60

📅 Oct 8, 2025 · 📍 Nationwide
Recalling company
Cellavision AB
Product
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.
Reason
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Oct 8, 2025
Reported
Nov 26, 2025
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
Status
Ongoing
Quantity
688 units
Lot / codes
UDI: 0735004097202/ Serial Numbers 63220-70093 or CN63220-CN70093 Serial Numbers 60001-63219 or CN60001-CN63219
Company location
Lund
Recall type
Voluntary: Firm initiated
Recall number
Z-0542-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0542-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.