Buy me a coffee
Home › FDA recalls › Brand Name: HeartMate Product Name: HeartMate 3 Controller
CLASS II · ModerateFDA

Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate 3 Controller

📅 Oct 9, 2025 · 📍 Nationwide
Recalling company
Thoratec LLC
Product
Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS
Reason
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Oct 9, 2025
Reported
Nov 26, 2025
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Status
Ongoing
Quantity
35376
Lot / codes
Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.
Company location
Pleasanton, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-0506-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0506-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.