Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload
- Recalling company
- Covidien, LP
- Product
- Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
- Reason
- A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 25, 2025
- Reported
- Nov 19, 2025
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
- Status
- Ongoing
- Quantity
- 3263 units
- Lot / codes
- CFN: SIGSDL45CTVT; UDI-DIs: (1) 20884521741840, (2) 10884521741843; Lot numbers: N5F1971Y, N5F1723Y, N5D1858UY, N5B1134UY;
- Company location
- North Haven, CT
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0485-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0485-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.