Medline Industries, LP recalls Convenience kits
- Recalling company
- Medline Industries, LP
- Product
- Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
- Reason
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Oct 10, 2025
- Reported
- Nov 19, 2025
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Status
- Ongoing
- Quantity
- 96,406 total units
- Lot / codes
- GYN ROBOTIC DYNJ901046M UDI-DI 40198459130381 (case), 10198459130380 (ea) Lots 25FBM810 25GBK307; ROBOTIC PROSTATE DYNJ909120D UDI-DI 40198459130541 (case), 10198459130540 (ea) Lots 25GBT821 25IBF809
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0450-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0450-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.