Medline Industries, LP recalls DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSC
- Recalling company
- Medline Industries, LP
- Product
- DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
- Reason
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Oct 10, 2025
- Reported
- Nov 19, 2025
- Distribution
- International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
- Status
- Ongoing
- Quantity
- 96,406 total units
- Lot / codes
- CYSTO DYNJ909127F UDI-DI 40198459055424 (case), 10198459055423 (ea) Lots 25FBP819 25GBM605; GENERAL LAPAROSCOPY-LF DYNJ909129D UDI-DI 40198459055264 (case), 10198459055263 (ea) Lots 25FBV537; ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909119G UDI-DI 40198459130527 (case), 10198459130526 (ea) Lots 25FBP681
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0446-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0446-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.