Buy me a coffee
Home › FDA recalls › MAGNETOM Cima.X Upgrade.
CLASS I · SeriousFDA

Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade.

📅 Oct 1, 2025 · 📍 CA
Recalling company
Siemens Medical Solutions USA, Inc
Product
MAGNETOM Cima.X Upgrade. Model Number: 11689304.
Reason
There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Oct 1, 2025
Reported
Nov 19, 2025
Distribution
US distribution to California. International distribution to Australia, Germany, Great Britian.
Status
Ongoing
Quantity
4 system (1 US; 3 OUS)
Lot / codes
Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.
Company location
Malvern, PA
Recall type
Voluntary: Firm initiated
Recall number
Z-0431-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0431-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.