Buy me a coffee
Home › FDA recalls › TruDi¿ Navigation System
CLASS II · ModerateFDA

Integra LifeSciences Corp. (NeuroSciences) recalls TruDi¿ Navigation System

📅 Oct 3, 2025 · 📍 Nationwide
Recalling company
Integra LifeSciences Corp. (NeuroSciences)
Product
TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
Reason
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Oct 3, 2025
Reported
Nov 12, 2025
Distribution
US Nationwide distribution.
Status
Ongoing
Quantity
1198 units
Lot / codes
Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;
Company location
Princeton, NJ
Recall type
Voluntary: Firm initiated
Recall number
Z-0420-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0420-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.