PHILIPS MEDICAL SYSTEMS recalls Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
- Recalling company
- PHILIPS MEDICAL SYSTEMS
- Product
- Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
- Reason
- Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 25, 2025
- Reported
- Nov 5, 2025
- Distribution
- Domestic: AL, AR, CA, CT, FL, GA, IL, IN, KY, MD, MI, OH, PA, PR, TN, TX, UT, VA; International: Argentina, Australia, Austria, Brazil, Czeck Republic, Denmark, Ecuador, Finland, France, Germany, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Kenya, Latvia, Lebanon, Libya, Netherlands, Nicaragua, Norway, Panama, Philippines, Poland, Portugal, Reunion, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, U.A.E., Yemen;
- Status
- Ongoing
- Quantity
- 165 units
- Lot / codes
- Product Code (REF): 728285; UDI-DI: 00884838113237; Serial Numbers: 800031, 800073, 800070, 35064, 35056, 35043, 35029, 800084, 800083, 800076, 800063, 800058, 800048, 800040, 800039, 800038, 800034, 800033, 800028, 800026, 800021, 800018, 800015, 800011, 800010, 800078, 800071, 35078, 35068, 35048, 35036, 800023, 800008, 800004, 800001, 35073, 35069, 35067, 35058, 35051, 35044, 35034, 35031, 35027, 35019, 35015, 35009, 35001, 800065, 35062, 35042, 35040, 35022, 35016, 35059, 800081, 800062, 800003, 35081, 35065, 35061, 35057, 35053, 35052, 35045, 35041, 35035, 35008, 800005, 35006, 800041, 35080, 800056, 35070, 35033, 35021, 800079, 800077, 800069, 800055, 800054, 800053, 800052, 800007, 800002, 35079, 35071, 35063, 35055, 35046, 35017, 35011, 35007, 35002, 800075, 800045, 800044, 800043, 800029, 800027, 800025, 800024, 800016, 800013, 800009, 800051, 800042, 800061, 800046, 800030, 800019, 800014, 800006, 800074, 800066, 35054, 35018, 800060, 800059, 800072, 35032, 35023, 35003, 800080, 800047, 35039, 35038, 35004, 76001, 35077, 35066, 35060, 35030, 35025, 35013, 35012, 35010, 35037, 35082, 35072, 35005, 35026, 800057, 800067, 800037, 800036, 800032, 800020, 35076, 35075, 35074, 35050, 35049, 35047, 35028, 35024, 35020, 35014, 800050, 800049, 800082, 800064, 800022, 800017, 800012;
- Company location
- Cambridge, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0376-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0376-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.