Boston Scientific Neuromodulation Corporation recalls Nalu TD2 Therapy Disc with firmware
- Recalling company
- Boston Scientific Neuromodulation Corporation
- Product
- Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System
- Reason
- Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 30, 2026
- Reported
- Nov 4, 2026
- Distribution
- US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
- Status
- Ongoing
- Quantity
- 134
- Lot / codes
- Firmware versions v2.0.71 and v2.0.77 REF/UDI-DI/Serial Numbers: 34014-001/00812537037134/ 100855, 100877, 100883, 100884, 100886, 100888, 100898, 100903, 100908, 100911, 100916, 100932, 101103, 100902; 74011/00812537037127, 00812537036977/101439, 100913, 100949, 101159, 101162, 101163, 101164, 101173, 101175, 101177, 101186, 101187, 101188, 101189, 101192, 101200, 101205, 101208, 101213, 101214, 101218, 101219, 101221, 101222, 101240, 101245, 101247, 101249, 101256, 101257, 101259, 101260, 101261, 101264, 101267, 101268, 101270, 101272, 101273, 101277, 101278, 101280, 101283, 101284, 101290, 101297, 101298, 101301, 101305, 101307, 101310, 101311, 101312, 101314, 101315, 101326, 101330, 101331, 101334, 101338, 101344, 101345, 101346, 101356, 101357, 101358, 101359, 101363, 101364, 101367, 101375, 101376, 101377, 101378, 101381, 101383, 101387, 101388, 101392, 101400, 101405, 101408, 101410, 101414, 101415, 101417, 101418, 101419, 101429, 101431, 101432, 101436, 101440, 101441, 101443, 101447, 101448, 101450, 101451, 101466, 101490, 101493, 101504, 101510, 101511, 101514, 101520, 101527, 101528, 101534, 101535, 101553, 101557, 101567, 101571, 101576, 101599, 101617, 101623, 101656
- Company location
- Valencia, CA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0311-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0311-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.