PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 5 M20.
- Recalling company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion 5 M20. Model Numbers: 722228, 722281.
- Reason
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Oct 6, 2026
- Reported
- Nov 4, 2026
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
- Status
- Ongoing
- Quantity
- 465 units
- Lot / codes
- 1. Model Number: 722228. UDI Numbers: (01)00884838099234(21)49, (01)00884838099234(21)159, (01)00884838099234(21)180, (01)00884838099234(21)273, (01)00884838099234(21)132, (01)00884838099234(21)533, (01)00884838099234(21)47, (01)00884838099234(21)522, (01)00884838099234(21)32, (01)00884838099234(21)23, (01)00884838099234(21)25, (01)00884838099234(21)62, (01)00884838099234(21)48, (01)00884838099234(21)153, (01)00884838099234(21)431, (01)00884838099234(21)113, (01)00884838099234(21)529, (01)00884838099234(21)659, (01)00884838099234(21)145, (01)00884838099234(21)567, (01)00884838099234(21)539, (01)00884838099234(21)645, (01)00884838099234(21)308, (01)00884838099234(21)333, (01)00884838099234(21)335, (01)00884838099234(21)570, (01)00884838099234(21)560, (01)00884838099234(21)44, (01)00884838099234(21)569, (01)00884838099234(21)418, (01)00884838099234(21)280, (01)00884838099234(21)616, (01)00884838099234(21)568, (01)00884838099234(21)471, (01)00884838099234(21)428, (01)00884838099234(21)465, (01)00884838099234(21)102, (01)00884838099234(21)176, (01)00884838099234(21)103, (01)00884838099234(21)178, (01)00884838099234(21)109, (01)00884838099234(21)16, (01)00884838099234(21)234, (01)00884838099234(21)566, (01)00884838099234(21)571, (01)00884838099234(21)463, (01)00884838099234(21)647, (01)00884838099234(21)317, (01)00884838099234(21)562, (01)00884838099234(21)552, (01)00884838099234(21)680, (01)00884838099234(21)724, (01)00884838099234(21)561, (01)00884838099234(21)517, (01)008848380
- Company location
- Best
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0295-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0295-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.