Buy me a coffee
Home › FDA recalls › Impella Optical Controller
CLASS I · SeriousFDA

Abiomed, Inc. recalls Impella Optical Controller

📅 Sep 18, 2026 · 📍 Nationwide
Recalling company
Abiomed, Inc.
Product
Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Reason
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Sep 18, 2026
Reported
Nov 4, 2026
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Status
Ongoing
Quantity
179 units
Lot / codes
Product Code: 0042-0010-US, 0042-0010-UK, 0042-0010-EU, 0042-0010-IN; UDI: 00813502010985, 00813502010978, 00813502010961, 00813502011425; Serial No. IC6962, IC7004, IC7036, IC7170, IC7213, IC7214, IC7239, IC7241, IC7677, IC7787, IC7880, IC8105, IC8106, IC8295, IC8308, IC8315, IC8316, IC8335, IC8541, IC8543, IC11456, IC7937, IC9832, IC9833, IC5409, IC7798, IC7800, IC7946, IC8070, IC8337, IC8441, IC7825, IC8126, IC8630, IC8640, IC9804, IC11451, IC11497, IC8388, IC8751, IC9501, IC9989, IC10102, IC10103, IC10318, IC10589, IC10463, IC10979, IC11299, IC8164, IC9348, IC9500, IC9581, IC9648, IC9777, IC9793, IC9797, IC9915, IC9919, IC9922, IC10278, IC10302, IC10983, IC10984, IC10985, IC11174, IC11410, IC13231, IC13263, IC13264, IC13265, IC13681, IC13682, IC10040, IC10043, IC8401, IC8748, IC8980, IC8609, IC9178, IC9177, IC9324, IC9328, IC9557, IC9558, IC9744, IC9745, IC9746, IC9908, IC9921, IC9969, IC9970, IC10063, IC10064, IC10149, IC10150, IC10238, IC10239, IC10341, IC10366, IC10367, IC12407, IC12408, IC12409, IC13057, IC13060, IC13061, IC13064, IC8333, IC8494, IC12326, IC4952, IC5041, IC5158, IC5281, IC5283, IC5284, IC5403, IC5404, IC5413, IC5527, IC5588, IC5635, IC5636, IC5641, IC5643, IC5666, IC5667, IC5804, IC5805, IC5918, IC5921, IC6013, IC6343, IC6421, IC6422, IC6461, IC6463, IC6465, IC6777, IC6785, IC6944, IC6945, IC6946, IC7161, IC7233, IC7232, IC7235, IC7626, IC7738, IC7739, IC7792, IC7793, IC7826, IC7956, IC8012, IC8258, IC8423, IC8424, IC8499, IC8511, IC8723, IC8596, IC86
Company location
Danvers, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-0209-2027
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0209-2027 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.