Abiomed, Inc. recalls Impella Optical Controller
- Recalling company
- Abiomed, Inc.
- Product
- Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
- Reason
- Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Sep 18, 2026
- Reported
- Nov 4, 2026
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
- Status
- Ongoing
- Quantity
- 179 units
- Lot / codes
- Product Code: 0042-0010-US, 0042-0010-UK, 0042-0010-EU, 0042-0010-IN; UDI: 00813502010985, 00813502010978, 00813502010961, 00813502011425; Serial No. IC6962, IC7004, IC7036, IC7170, IC7213, IC7214, IC7239, IC7241, IC7677, IC7787, IC7880, IC8105, IC8106, IC8295, IC8308, IC8315, IC8316, IC8335, IC8541, IC8543, IC11456, IC7937, IC9832, IC9833, IC5409, IC7798, IC7800, IC7946, IC8070, IC8337, IC8441, IC7825, IC8126, IC8630, IC8640, IC9804, IC11451, IC11497, IC8388, IC8751, IC9501, IC9989, IC10102, IC10103, IC10318, IC10589, IC10463, IC10979, IC11299, IC8164, IC9348, IC9500, IC9581, IC9648, IC9777, IC9793, IC9797, IC9915, IC9919, IC9922, IC10278, IC10302, IC10983, IC10984, IC10985, IC11174, IC11410, IC13231, IC13263, IC13264, IC13265, IC13681, IC13682, IC10040, IC10043, IC8401, IC8748, IC8980, IC8609, IC9178, IC9177, IC9324, IC9328, IC9557, IC9558, IC9744, IC9745, IC9746, IC9908, IC9921, IC9969, IC9970, IC10063, IC10064, IC10149, IC10150, IC10238, IC10239, IC10341, IC10366, IC10367, IC12407, IC12408, IC12409, IC13057, IC13060, IC13061, IC13064, IC8333, IC8494, IC12326, IC4952, IC5041, IC5158, IC5281, IC5283, IC5284, IC5403, IC5404, IC5413, IC5527, IC5588, IC5635, IC5636, IC5641, IC5643, IC5666, IC5667, IC5804, IC5805, IC5918, IC5921, IC6013, IC6343, IC6421, IC6422, IC6461, IC6463, IC6465, IC6777, IC6785, IC6944, IC6945, IC6946, IC7161, IC7233, IC7232, IC7235, IC7626, IC7738, IC7739, IC7792, IC7793, IC7826, IC7956, IC8012, IC8258, IC8423, IC8424, IC8499, IC8511, IC8723, IC8596, IC86
- Company location
- Danvers, MA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0209-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0209-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.