Beckman Coulter, Inc. recalls ROTOR ASSY-SW41
- Recalling company
- Beckman Coulter, Inc.
- Product
- ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
- Reason
- Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 21, 2026
- Reported
- Oct 28, 2026
- Distribution
- US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 336 units
- Lot / codes
- Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Company location
- Indianapolis, IN
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0202-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0202-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.