Medline Industries, LP recalls Medline surgical convenience kits labeled as: 1) EAR
- Recalling company
- Medline Industries, LP
- Product
- Medline surgical convenience kits labeled as: 1) EAR, Medline product number/SKU DYNJ909020B; 2) EAR, Medline product number/SKU DYNJ909020A; 3) CAROTID PACK, Medline product number/SKU DYNJ42922C; 4) EAR, Medline product number/SKU DYNJ909020; 5) FEM CAROTID PACK, Medline product number/SKU DYNJ65706; 6) CAROTID CDS-LF, Medline product number/SKU CDS983139B.
- Reason
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 2, 2026
- Reported
- Oct 21, 2026
- Distribution
- US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 620 kits
- Lot / codes
- Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 26AMF311; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25JMD500; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25DMJ849; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25CMD823; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 25BMG250; Medline product number/SKU DYNJ909020B, UDI/DI 10198459159213 (each) 40198459159214 (case), Lot number 24LMG633; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24KMB736; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24JMD634; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24IMD605; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 24HBT243; Medline product number/SKU DYNJ42922C, UDI/DI 10193489335019 (each) 40193489335010 (case), Lot number 24HBD143; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24GMG103; Medline product number/SKU DYNJ909020A, UDI/DI 10195327665760 (each) 40195327665761 (case), Lot number 24DMI379; Medline product numb
- Company location
- Northfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0173-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0173-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.