IMPRIS Imaging Inc recalls IMRIS Operating Suite
- Recalling company
- IMPRIS Imaging Inc
- Product
- IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
- Reason
- There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Sep 30, 2025
- Reported
- Nov 5, 2025
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
- Status
- Ongoing
- Quantity
- N/A
- Lot / codes
- IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.
- Company location
- Chaska, MN
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0162-2026
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0162-2026 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.