CLASS I · SeriousFDA

IMPRIS Imaging Inc recalls IMRIS Operating Suite

📅 Sep 30, 2025 · 📍 Nationwide
Recalling company
IMPRIS Imaging Inc
Product
IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
Reason
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Sep 30, 2025
Reported
Nov 5, 2025
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Status
Ongoing
Quantity
N/A
Lot / codes
IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.
Company location
Chaska, MN
Recall type
Voluntary: Firm initiated
Recall number
Z-0162-2026
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0162-2026 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.