Philips Ultrasound, LLC recalls Philips Lumify Diagnostic Ultrasound System
- Recalling company
- Philips Ultrasound, LLC
- Product
- Philips Lumify Diagnostic Ultrasound System
- Reason
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- Classification
- Class I – a reasonable probability that use will cause serious adverse health consequences or death
- Recall date
- Sep 24, 2026
- Reported
- Oct 28, 2026
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Status
- Ongoing
- Quantity
- 13,206
- Lot / codes
- UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X
- Company location
- Bothell, WA
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0152-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0152-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.