Baxter Healthcare Corporation recalls Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Pro
- Recalling company
- Baxter Healthcare Corporation
- Product
- Likorall Overhead Lifts Product Code: 3121001 Product Description: LIKORALL 200 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122008NA Product Description: LIKORALL 242 ES R2R, WHITE Product Code: 3122009 Product Description: LIKORALL 242 S, NATURAL Product Code: 3122009NA Product Description: LIKORALL 242 S, NATURAL Product Code: 3122010 Product Description: LIKORALL 242 S, WHITE Product Code: 3122011 Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122011NA Product Description: LIKORALL 242 S R2R, NATURAL Product Code: 3122012 Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122012NA Product Description: LIKORALL 242 S R2R, WHITE Product Code: 3122501 Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122501NA Product Description: LIKORALL 250 ES, NATURAL Product Code: 3122502 Product Description: LIKORALL 250 ES, WHITE Product Code: 3122502NA Product Description: LIKORALL 250 ES, WHITE Product Code: 3123001 Product Description: LIKORALL 243 ES Product Code: 3123001NA Product Description: LIKORALL 243 ES Product Code: 3123002 Product Description: LIKORALL 243 ES, WHITE Product Code: 3123002NA Product Description: LIKORALL 243 ES, WHITE Product Code: 3124050 Product Description: LIKORALL 250 S, IRC
- Reason
- Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Sep 14, 2026
- Reported
- Oct 21, 2026
- Distribution
- Worldwide - US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, and Wyoming. International distribution in pending.
- Status
- Ongoing
- Lot / codes
- Product Code: 3121001 Product Description: LIKORALL 200 Lot/Serial Number: Intervals: 8400054-8450499 and 9000000-9000999 UDI-DI Number: 00887761981911 Product Code: 3122005 Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Greater than or equal to 8104548 UDI-DI Number: 00887761979710 Product Code: 3122005NA Product Description: LIKORALL 242 ES, NATURAL Lot/Serial Number: Interval: 429087-431032 UDI-DI Number: 00887761979727 Product Code: 3122006 Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Greater than or equal to: 8101715 UDI-DI Number: 00887761979734 Product Code: 3122006NA Product Description: LIKORALL 242 ES, WHITE Lot/Serial Number: Interval: 429186-431035 UDI-DI Number: 00887761979741 Product Code: 3122007 Product Description: LIKORALL 242 ES R2R, NATURAL Lot/Serial Number: Greater than or equal to: 8104481 UDI-DI Number: 00887761979758 Product Code: 3122007NA Product Description: LIKORALL 242 ES R2R, NATURAL Lot/Serial Number: Interval: 429041-430984 UDI-DI Number: 00887761979765 Product Code: 3122008 Product Description: LIKORALL 242 ES R2R, WHITE Lot/Serial Number: Greater than or equal to: 8100815 UDI-DI Number: 00887761979772 Product Code: 3122008NA Product Description: LIKORALL 242 ES R2R, WHITE Lot/Serial Number: Interval: 429081-431018 UDI-DI Number: 00887761979789 Product Code: 3122009 Product Description: LIKORALL 242 S, NATURAL Lot/Serial Number: Greater than or equal to: 8000235 UDI-DI Number: 0088776197
- Company location
- Deerfield, IL
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0151-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0151-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.