Home › FDA recalls › Product Name: Nihon Kohden CNS-6200 Series Central Nurse Sta
CLASS II · ModerateFDA

Nihon Kohden America LLC recalls Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station

📅 Sep 8, 2026 · 📍 Nationwide
Recalling company
Nihon Kohden America LLC
Product
Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No
Reason
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Classification
Class II – may cause temporary or medically reversible health consequences
Recall date
Sep 8, 2026
Reported
Oct 14, 2026
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
Status
Ongoing
Quantity
808
Lot / codes
Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
Company location
Irvine, CA
Recall type
Voluntary: Firm initiated
Recall number
Z-0089-2027
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0089-2027 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.