Steris Corporation recalls VERIFY Biological Steam Test Pack
- Recalling company
- Steris Corporation
- Product
- VERIFY Biological Steam Test Pack Model/Catalog Number: S3069
- Reason
- Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
- Classification
- Class II – may cause temporary or medically reversible health consequences
- Recall date
- Aug 31, 2026
- Reported
- Oct 14, 2026
- Distribution
- Worldwide - US Nationwide distribution.
- Status
- Ongoing
- Quantity
- 233
- Lot / codes
- UDI:10724995023864 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
- Company location
- Mentor, OH
- Recall type
- Voluntary: Firm initiated
- Recall number
- Z-0075-2027
- Agency
- U.S. Food and Drug Administration (FDA)
View official FDA notice ↗ Search recall number Z-0075-2027 in the FDA Enforcement Report. Also see FDA recall press releases.
If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.