Instrumentation Laboratory recalls HemoslL AcuStar HIT-lgG

📅 Sep 3, 2026 · 📍 Nationwide
Recalling company
Instrumentation Laboratory
Product
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnostic
Reason
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Classification
Class I – a reasonable probability that use will cause serious adverse health consequences or death
Recall date
Sep 3, 2026
Reported
Oct 21, 2026
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.
Status
Ongoing
Quantity
434 units
Lot / codes
Part Number: 0009800127. UDI-DI: 08426951126545. Lot Number (Expiration Date): B40005 (01/16/2027), B38062 (11/01/2026).
Company location
Bedford, MA
Recall type
Voluntary: Firm initiated
Recall number
Z-0001-2027
Agency
U.S. Food and Drug Administration (FDA)

View official FDA notice ↗ Search recall number Z-0001-2027 in the FDA Enforcement Report. Also see FDA recall press releases.

If you use this product, check the lot, NDC or model number against the official notice and follow its instructions. Do not stop a prescribed medicine or remove a medical device without talking to your doctor or pharmacist. Report problems to FDA MedWatch. FoodRecallNow summarizes public FDA data and is not medical advice.